Calibration of Orifice Plates            

Theory

Plate of orifice, containing a concentric circular cut, place into duct such that the plate of orifice is concentric with the duct (Figure1). Air quantity following through the duct with the presence of plate of orifice is formulated as

$$ Q= \alpha A_2 \sqrt{\frac{2(\delta P)}{\rho - (1-m^2)}} -------------------- (1) $$

Where, α = coefficient of discharge
A2 = Area of cross-section of the opening of the orifice plate
ΔP = Static pressure difference of the two sides of orifice plate i.e. (P1−P2). In this case, P1 is the static pressure before the plate and P2 is the static pressure after the plate.
ρ = Density of medium; in this case the medium is air so its value is 1.
m = Cross-section area ratio of plate of orifice opening with the duct opening = A2/A1 where as A1 is the cross-section area of the opening of the duct. In this experiment the design is in such a way that the value of m is 0.5.
In this experiment, the discharge coefficient will be determined from the equation1 with varying the air quantity by changing the throttling device at the end of the duct and calculating pressure difference by measuring pressures at the two sides of the plate of orifice. Air quantity will be determined by multiplying the area with air velocity measured by the anemometer. So from the equation1 the discharge coefficient α will be determined which is given below in equation2.

$$ \alpha = \frac{Q}{A_2 \sqrt{\frac{2(\delta P)}{\rho - (1-m^2)}}} ----------------- (2) $$

Instruments:

The required instruments in this experiment are given below.

  1. Orifice plate (Situated into the duct)
  2. Anemometer
  3. Throttling devices
  4. U tube manometer
  5. Measuring tape

Demo